• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CARDIOVASCULAR SYSTEMS INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE); CORONARY ATHERECTOMY DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CARDIOVASCULAR SYSTEMS INC. DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE); CORONARY ATHERECTOMY DEVICE Back to Search Results
Model Number GWC-12325LG-FT
Device Problems Unintended System Motion (1430); Material Separation (1562)
Patient Problem Foreign Body In Patient (2687)
Event Date 01/20/2024
Event Type  Injury  
Manufacturer Narrative
D10: unknown part/model number and unknown lot number.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The device history record for the reported oad could not be reviewed as the lot number was not provided.If the lot number is provided, a dhr review will be performed.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: 13887 and 13911.
 
Event Description
During the treatment of a 90% stenosed lesion in left anterior descending (lad) vessel, while the orbital atherectomy system (oas) was being used at low speed in the pushing direction, the crown jumped in the spring tip section of the viperwire advance guidewire which led to spring tip fracture.A non-abbott micro snare was used to retrieve the fractured viperwire guidewire from the patient's body.Percutaneous coronary intervention (pci) was completed successfully without any complications to the patient.After retrieval of the fractured viperwire, the procedure was completed with a balloon and a stent.The fractured viperwire was discarded.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DIAMONDBACK 360 CORONARY ORBITAL ATHERECTOMY SYSTEM (VIPERWIRE ADVANCE)
Type of Device
CORONARY ATHERECTOMY DEVICE
Manufacturer (Section D)
CARDIOVASCULAR SYSTEMS INC.
1225 old highway 8 nw
st. paul
Manufacturer (Section G)
CARDIOVASCULAR SYSTEMS INC.
1225 old highway 8 nw
st. paul 55112
Manufacturer Contact
keerthi bangalore ramanath
1225 old highway 8 nw
st. paul 55112
MDR Report Key18726919
MDR Text Key335709793
Report Number3004742232-2024-00107
Device Sequence Number1
Product Code MCX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P130005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberGWC-12325LG-FT
Device Catalogue Number7-10038-02
Device Lot Number470337-1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/22/2024
Initial Date FDA Received02/16/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured01/18/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
UNKNOWN CORONARY OAD
Patient Outcome(s) Required Intervention;
-
-