D10: unknown part/model number and unknown lot number.The material inspection report for this guide wire lot number has been reviewed.No issues or discrepancies were noted during this review that would have contributed to the reported event.The device met material, assembly, and quality control requirements.The device history record for the reported oad could not be reviewed as the lot number was not provided.If the lot number is provided, a dhr review will be performed.The results of the investigation are inconclusive since the reported device was not returned for analysis.Based on the information received, the cause of the reported event could not be conclusively determined.Csi id: 13887 and 13911.
|
During the treatment of a 90% stenosed lesion in left anterior descending (lad) vessel, while the orbital atherectomy system (oas) was being used at low speed in the pushing direction, the crown jumped in the spring tip section of the viperwire advance guidewire which led to spring tip fracture.A non-abbott micro snare was used to retrieve the fractured viperwire guidewire from the patient's body.Percutaneous coronary intervention (pci) was completed successfully without any complications to the patient.After retrieval of the fractured viperwire, the procedure was completed with a balloon and a stent.The fractured viperwire was discarded.
|