MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Model Number AFAPRO28 |
Device Problem
Adverse Event Without Identified Device or Use Problem (2993)
|
Patient Problems
Low Blood Pressure/ Hypotension (1914); Pericardial Effusion (3271)
|
Event Date 02/14/2024 |
Event Type
Injury
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during a cryo ablation procedure, there was an acute drop in the patient's blood pressure.Intracardiac echocar diography (ice) was utilized and pericardial effusion was observed.The procedure was aborted while the patient was under general anesthesia.A pericardiocentesis was performed, but the patient became increasingly hemodynamically unstable.An emergency sternotomy and surgical repair were subsequently performed.The patient was stabilized and transferred to the intensive care unit (icu).No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Continuation of d10: product id:990063-020 product type: mapping catheter, product id: 4fc12 product type: sheath.Product event summary: the data file was returned and analyzed.One patient file was returned and analysed.The patient data file showed ten applications were performed using a console of serial number (b)(6) and a balloon catheter identified as 2af284 of lot number 10289.The patient file did not show any system notice on the reported date of the event.The failure file was not received.In conclusion, the decision to abort the procedure without use of alternate therapy was based upon the medical judgment of the physician.Additionally, the reported pericardial effusion and hypotension occurred during the procedure and could not be confirmed through data analysis.There is no indication of a relation of the adverse event to the performance and malfunction of the product.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|