• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION SYNERGY XD; CORONARY DRUG-ELUTING STENT Back to Search Results
Lot Number 0030214117
Device Problems Device Dislodged or Dislocated (2923); Material Integrity Problem (2978)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/27/2023
Event Type  malfunction  
Event Description
It was reported that stent dislodgement occurred.The 75% stenosed target lesion was located in the moderately tortuous and moderately calcified left anterior descending.A 3.50 x 12mm synergy xd drug-eluting stent was advanced for treatment.However, during insertion of the device the stent dislodged inside the patient.The procedure was completed with a different device.There were no patient complications nor injuries reported.
 
Manufacturer Narrative
Device evaluated by mfr.Synergy xd mr ous 3.50 x 12mm stent delivery system (sds); catheter number was returned for analysis.A visual and tactile examination of the hypotube found multiple kinks along the shaft.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues along the shaft polymer extrusion.No stent was returned for analysis.The balloon cones were reviewed, and the balloon was in a deflated state.Based on microscopic analysis, no stent was returned for analysis.A review of the manufacturing stent profile data was performed and the stent od at the time of manufactured.The balloon cones were reviewed, and the balloon was in a deflated state.A visual and tactile examination of the hypotube found multiple kinks along the shaft.Bumper tip showed no signs of distal tip damage.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYNERGY XD
Type of Device
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
rachel shields
4100 hamline ave n
arden hills, MN 55112
6512422111
MDR Report Key18727411
MDR Text Key336192443
Report Number2124215-2024-08075
Device Sequence Number1
Product Code NIQ
Combination Product (y/n)Y
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,Company Representative,Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Lot Number0030214117
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/16/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/24/2024
Initial Date FDA Received02/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/15/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-