The device was returned, and the evaluation found the instrument channel port was loose.After re-installation, the fault was eliminated.Should additional relevant information become available, a supplemental report will be submitted.
|
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the loose biopsy valve was excessive force applied to the biopsy channel port when the user detached forceps/irrigation plug from the port.The event can be detected/prevented by following the instructions for use which state: ¿·operation manual_ preparation and inspection_ inspection of the endoscope inspect the external surface of the entire insertion tube including the bending section and the distal end for dents, bulges, swelling, peeling, scratching, holes, sagging, transformation, bends, adhesion of foreign body, dropout of parts, any protruding objects or other irregularities.¿ olympus will continue to monitor field performance for this device.
|