Model Number US5062
Device Problems Fluid/Blood Leak (1250); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Insufficient Information (4580)
Event Date 01/10/2024
Event Type  Death  
Manufacturer Narrative
A2 - patient age requested but not yet provided.No further information was provided.A supplemental report will be submitted once the manufacturer¿s investigation is complete.
 
Event Description
It was reported that after the patient was initiated on extracorporeal membrane oxygenation (ecmo) support, blood was noted to be leaking out of the bottom of the oxygenator, which was noted to be very alarming.The leak was immediately noticed as soon as blood started to flow through the oxygenator, it also started to come out the bottom exhaust as well.The pump speed was 2400 rpm, the blood flow was 4.3 l/min, and the patient's activated clotting time (act) was 210.The oxygenator was exchanged.The patient lost blood that was in the oxygenator when it was removed.It was noted that the oxygenator could possibly contain biological residue of who risk group 4 or 3 such as avian flu, sars-cov-2 etc.Which may be infectious, however is was unknown; the device did not contain biological residue of who risk group 1 or 2.Additional information was provided that the patient passed away.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADVANCED MEMBRANE GAS EXCHANGE PMP STERILE
Type of Device
OXYGENATOR, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
EUROSETS S.R.L.
strada statale 12, n°143
medolla, modena 41036
IT  41036
Manufacturer (Section G)
EUROSETS S.R.L.
strada statale 12, n°143
medolla, modena 41036
IT   41036
Manufacturer Contact
lindsey sallese
6035 stoneridge drive
pleasanton, CA 94588
7818528207
MDR Report Key18731321
MDR Text Key335692458
Report Number3003752502-2024-00004
Device Sequence Number1
Product Code DTZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K141492
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/19/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUS5062
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/01/2024
Initial Date FDA Received02/19/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Death;
Patient SexMale
Patient Weight89 KG
-
-