Model Number CLV-190
Device Problem Loss of Power (1475)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned for evaluation and the evaluation was unable to duplicate the reported issue.Testing and inspection of the device found the lamp bulb had been used for over 500 hours and reached the end of its life.The investigation is ongoing, and a supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
It was reported, the light source would completely shut off by itself.The team would have to reactivate the unit by pressing the off/on button as if they were turning it on for the first time for the unit to function again.The issue occurred during an unspecified diagnostic procedure.There were no reports of patient harm or impact associated with this event.
 
Manufacturer Narrative
This supplemental report is being submitted to provide the correction of the initial mdr and the legal manufacturer's final investigation.Corrected fields: b5 and e3.Additional information: h6.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over eight (8) years since the subject device was manufactured.A definitive root cause was not identified, however based on the results of the investigation, the probable cause of the "shuts down" malfunction could not be determined.Olympus will continue to monitor field performance for this device.
 
Event Description
The costumer reported that the issue occurred prior the procedure started, and it was completed without delay, with same set of equipment.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA III XENON LIGHT SOURCE
Type of Device
XENON LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18731591
MDR Text Key336636620
Report Number3002808148-2024-01508
Device Sequence Number1
Product Code NWB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131780
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLV-190
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/25/2024
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/19/2024
Supplement Dates Manufacturer Received03/01/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-