Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/07/2024
Event Type  malfunction  
Event Description
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor auto zero failed error code.The device was reported to be outside of use at the time of the reported problem.There was no patient or user harm reported.The authorized service provider (asp) confirmed the error code in the device event log.The asp replaced the flow sensor assembly (fsa) to resolve the proximal pressure sensor auto zero failed error code.The asp completed a comprehensive inspection, cleaning, running test, and function test following the repair.No other abnormality was observed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18735523
MDR Text Key335709581
Report Number2518422-2024-08328
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1076709
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/07/2024
Initial Date FDA Received02/19/2024
Date Device Manufactured02/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-