RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/07/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor auto zero failed error code.The device was reported to be outside of use at the time of the reported problem.There was no patient or user harm reported.The authorized service provider (asp) confirmed the error code in the device event log.The asp replaced the flow sensor assembly (fsa) to resolve the proximal pressure sensor auto zero failed error code.The asp completed a comprehensive inspection, cleaning, running test, and function test following the repair.No other abnormality was observed.
|
|