C.R. BARD, INC. (BASD) -3006260740 FULL KIT 18G X 8CM WITH PROBE COVER; CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS Back to Search Results
Model Number N/A
Device Problem Break (1069)
Patient Problem Foreign Body In Patient (2687)
Event Date 12/14/2023
Event Type  Injury  
Event Description
It was reported by customer, "midline was noticed to be broken off at the insertion site.The patient required x-rays to confirm the site of retained catheter fragments and surgically removed.Fragments measured 9cm." no other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The device has not been returned to the manufacturer for evaluation.H3 other text : device not returned for evaluation.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The complaint of a break in the catheter was confirmed.The product returned for evaluation was one catheter hub with a small portion of catheter from an 18 ga powerglide pro.The investigation findings are consistent with catheter damage caused by contact with a sharp edged instrument such as a scalpel.The returned product sample was evaluated and a break was observed just distal of the luer.Microscopic examination of the catheter break confirmed it was typical of contact with a sharp bladed instrument, and the characteristics observed which supported this type of failure included: the fracture surface was reflective in nature and contained a striated pattern.This can occur due to pattern transfer of the sharpened edge typically found on a scalpel-type instrument.Sharply formed fracture edges planar shape to the fracture surface an examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.H3 other text : evaluation findings are in section h.11.
 
Event Description
It was reported via medwatch, "midline was noticed to be broken off at the insertion site.The patient required x-rays to confirm the site of retained catheter fragments and surgically removed.Fragments measured 9cm." no other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FULL KIT 18G X 8CM WITH PROBE COVER
Type of Device
CATHETER,INTRAVASCULAR,THERAPEUTIC,SHORT-TERM LESS THAN 30 DAYS
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
johanna de oliveira
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18740151
MDR Text Key335768152
Report Number3006260740-2024-00677
Device Sequence Number1
Product Code FOZ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K162377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/17/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue NumberF218088PT
Device Lot NumberREHR0753
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/26/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received04/17/2024
Supplement Dates FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-