E1: initial reporter last name: (b)(6).E1: initial reporter address: (b)(6).H4: the device was manufactured between 10may2023 and 11may2023.H10: the device was received for evaluation not in its original unopened baxter package.Visual inspection noted the fill pot cap was missing from the infusor sample.The reported condition was verified.As the device was not received in the original unopened packaging, the cause of the condition could not be determined.Although, during manufacturing the device was 100% visually inspected for missing components; therefore, a missing fill port cap would have been detected.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|