RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint on the v60 ventilator indicating that there was a proximal pressure sensor auto zero failed alarm.The device was reported to be outside of use at the time of the reported problem.No patient or user harm reported.The remote service engineer (rse) informed the customer to perform the following troubleshooting steps in this order: verify the pin alignment between the solenoids and data acquisition (da) printed circuit board assembly (pcba), replace the 3rd and 4th proximal autozero solenoids, and then replace the da pcba.The customer requested the part information for the solenoids and da pcba and the rse provided the part numbers for both.
|
|
Manufacturer Narrative
|
Multiple good faith efforts (gfe) were attempted to obtain confirmation of the part order, part replacement, and resolution.No response or further information was received from the customer.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
|
|
Search Alerts/Recalls
|
|
|