This report is being supplemented to provide additional information based on the legal manufacturer's investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the piece chipped off was unknown.The event can be detected/prevented by following the instructions for use which state: ¿caution do not coil the insertion tube, universal cord, or video cable with a diameter of less than 10 cm.Equipment damage may result.Do not attempt to bend or twist the endoscope¿s insertion section with excessive force.The insertion section may be damaged.Do not apply shock to the distal end of the insertion section, particularly the objective lens surface at the distal end.Visual abnormalities may result.If the endoscope is dropped or the distal end of the endoscope receives a hard impact, the endoscope may be damaged even if no visible damage of the lens on the distal end can be found.In this case, stop using the endoscope, and contact olympus.Do not twist or bend the bending section with your hands.Equipment damage may result.Do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks.¿ olympus will continue to monitor field performance for this device.
|