• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; REPLACEMENT HEART VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; REPLACEMENT HEART VALVE Back to Search Results
Model Number 8300AB
Device Problems Perivalvular Leak (1457); Insufficient Information (3190)
Patient Problems Dyspnea (1816); Heart Failure/Congestive Heart Failure (4446); Insufficient Information (4580)
Event Date 01/25/2024
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress; therefore, a conclusion has yet to be established.A supplemental report will be submitted accordingly upon investigation completion.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
Through implant patient registry it was learned that a 23mm 8300ab valve was explanted after an implant duration of 7 (seven) years, 2 (two) months due to unknown reasons.The explanted valve was replaced with a 25mm 11500a valve.Edwards lifesciences maintains an implant patient registry.This registry is a patient tracking mechanism for serialized edwards implantable devices (bioprosthetic heart valves and annuloplasty rings), rather than a true post-market surveillance registry.Through the registry, edwards is notified when these devices are implanted.In addition, patient and/or device status may be reported to the registry via the implantation data cards.The information is received from various sources (e.G.Surgeon, hospital, and patient family members) and is not received in the form of a conventional 'customer complaint'.The information reported may or may not be related to the edwards device.
 
Event Description
Through implant patient registry it was learned that a 23mm 8300ab valve was explanted after an implant duration of 7 (seven) years, 2 (two) months due to paravalvular leak.The patient presented with heart failure and shortness of breath.The explanted valve was replaced with a 25mm 11500a valve.The patient was stable post procedure.
 
Manufacturer Narrative
The device history record (dhr) review was completed, and this device passed all manufacturing and sterilization inspections prior to release for distribution.There were no issues identified that would have impacted this event.Regurgitation is considered to be a perivalvular leak (pvl) if a turbulent eccentric jet originates between the bioprosthetic sewing ring and the annulus.Pvl can occur in the mitral and aortic position for similar reasons.In the early postoperative period, the highest incidence of pvl has been seen in patients developing infective endocarditis, which is most likely attributed to inadequate peri-operative antibiotic prophylaxis or nosocomial infection.Annular calcification is also a risk factor for the development of peri-operative pvl as the bioprosthesis may not seat properly after debridement.Technique related factors, such as incorrect valve sizing, have been shown to contribute to the development of pvl.Anatomical factors may create difficulty seating the bioprosthetic valve resulting in pvl.The anatomy of the annulus may induce mechanical stresses along the rigid bioprosthetic ring which can influence long-term valve performance and durability.A diseased or rigid annulus can potentially increase the mechanical stress on the prosthetic valve, leading to pvl.A definitive root cause cannot be conclusively determined; however, patient and/or procedural factors likely caused or contributed.Edwards will continue to review and monitor all reported events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
REPLACEMENT HEART VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES LLC
1 edwards way
irvine CA 92614
Manufacturer Contact
saurav singh
1 edwards way
mle fl2 - office m2013
irvine, CA 92614
9492506615
MDR Report Key18744258
MDR Text Key335813347
Report Number2015691-2024-01181
Device Sequence Number1
Product Code DYE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date 04/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/21/2017
Device Model Number8300AB
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/06/2024
Initial Date FDA Received02/20/2024
Supplement Dates Manufacturer Received02/22/2024
04/01/2024
Supplement Dates FDA Received03/04/2024
04/16/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Disability; Life Threatening;
Patient Age65 YR
Patient SexMale
-
-