This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.The device history record was unable to be reviewed for this device since the lot number was not provided.However, olympus only releases products to market that meet all manufacturing specifications and final product release criteria.Based on the results of the investigation, external stress was applied to lock lever of the connecting tube, which led to breakage of the tube.The cause of the external stress on the lock lever was not established.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿visually inspect the connecting tube to ensure that there are no cracks, breaks, rips, scratches, attached debris, or other irregularities.Move the lock levers of the reprocessor side connector to make sure that they function properly and are not broken.¿ olympus will continue to monitor field performance for this device.
|