MEDTRONIC PUERTO RICO OPERATIONS CO. INTERSTIM X; STIMULATOR, ELECTRICAL, IMPLANTABLE, FOR INCON
|
Back to Search Results |
|
Model Number 97800 |
Device Problem
Pocket Stimulation (1463)
|
Patient Problems
Undesired Nerve Stimulation (1980); Urinary Frequency (2275)
|
Event Date 05/16/2023 |
Event Type
malfunction
|
Event Description
|
Information was received from a patient who was implanted with an implantable neurostimulator (ins) for urinary dysfunction/sacral nerve stim.The reason for call was patient said they are still using the restroom a lot at night.Patient said they didn't know they could change programs.Agent did not ask about the circumstances that led to the reported issue.Reviewed how to switch programs.Patient was able to change programs and increase stim.Patient was reporting stim in their back at the ins site but kept increasing and was able to feel stim in the bike seat area.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Event Description
|
Additional information was received from the patient.They reported that the cause of feeling stim in the back at the ins site when increasing stimulation was not determined.A likely cause was noted as "no one has ever talked with me about changing the settings or what to do if the program isn't working or helping me.A lack of education from office staff." the steps taken were noted as being "i will keep try the different programs to try to get a better experience/good night rest." the issue has not yet been resolved.
|
|