WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 02/07/2024
Event Type  Injury  
Manufacturer Narrative
Manufacturers ref# (b)(4).Blank fields on this form indicate the information is unknown or unavailable.G4) pma/510(k): k211875.Investigation is still in progress.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Description of event according to initial reporter: fractured cook celect inferior vena cava filter.Required intervention to prevent permanent impairment/damage.
 
Manufacturer Narrative
Manufacturer ref#: (b)(4).H11) corrected data compared with medwatch report: 5151583.D1) celect inferior vena cava filter.D3) cook medical llc.E1) (b)(4).Summary of investigational findings: fractured cook celect inferior vena cava filter.Required intervention to prevent permanent impairment/damage.The fractured celect filter was not returned and based on the information provided it would be inappropriate to speculate at what may or may not have caused the filter to fracture during an implant period of more than ten years.Filter fracture has been reported and may be either symptomatic or asymptomatic.Fracture of a filter leg may be due to repetitive motion on a filter leg in an unusual, stressed position, such as a filter leg penetrating/perforating the ivc; or a filter leg being caught in a side branch (e.G., a renal vein).Other potential causes of filter fracture may include excessive force or manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Retrieval of a fractured filter or filter fragments (including embolized fragments) using endovascular techniques has been reported.Potential adverse events that may occur include, but are not limited to, the following: filter fracture, filter or filter fragment embolization, trauma to adjacent structures.There are adequate controls in place to ensure that the type of device was manufactured to specifications.Cook was unable to conduct a review of the device history record, as the lot number of the complaint device was not provided for the investigation.Cook medical will continue to monitor for similar events.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned; that a death or serious injury occurred; or that any cook device caused or contributed to; or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Medwatch: 5151583 received 07mar2024.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA   4632
Manufacturer Contact
alex rahbek
sandet 6
bjaeverskov 4632
DA   4632
56868686
MDR Report Key18745659
MDR Text Key335809842
Report Number3002808486-2024-00035
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 03/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/08/2024
Initial Date FDA Received02/21/2024
Supplement Dates Manufacturer Received03/06/2024
Supplement Dates FDA Received03/13/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age71 YR
Patient SexFemale
Patient Weight100 KG
-
-