• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SENSEONICS INC. EVERSENSE SENSOR; EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM Back to Search Results
Model Number 101368-67A
Device Problem Improper or Incorrect Procedure or Method (2017)
Patient Problem Pain (1994)
Event Date 07/22/2019
Event Type  Injury  
Manufacturer Narrative
This report is being submitted retrospectively as part of internal review.It was discovered that the sensor was placed directly on a nerve and the pain caused by the sensor immediately subsided after its removal.The sensor was remove dfrom the user's arm.He pain caused by the sensor immediately subsided after the removal.The user is doing well.
 
Event Description
Senseonics became aware of an incident where the patient was in considerable pain, he could not move his arm properly, nor could he feel or find the sensor when using the placement guide.User said that he was in chronic pain and is very sensitive.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18747314
MDR Text Key335871231
Report Number3009862700-2024-00113
Device Sequence Number1
Product Code QCD
Combination Product (y/n)Y
Reporter Country CodeGM
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date09/07/2019
Device Model Number101368-67A
Device Catalogue NumberFG-4400-XX-302
Device Lot NumberWP04711
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 11/11/2019
Initial Date FDA Received02/21/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured09/10/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-