A.1.-a.5.There was no known patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa, and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in france.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.A livanova field service engeneer was dispatched the customer and the device was disassembled.The following components were replaced: processor board cpu ul508, water drain and sniffle valves, circulator motor em109, patient and cardioplegia pumps.The motor was found noisy motors, temperature adjustments.
|