Model Number 16-02-80
Device Problem Pumping Problem (3016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Manufacturer Narrative
A.1.-a.5.There was no known patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa, and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in france.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.A livanova field service engeneer was dispatched the customer and the device was disassembled.The following components were replaced: processor board cpu ul508, water drain and sniffle valves, circulator motor em109, patient and cardioplegia pumps.The motor was found noisy motors, temperature adjustments.
 
Event Description
Livanova deutschland received a report that a heater-cooler system 3t displayed an error message pump 1 uses too much power on cardioplegia circuit (e06) error pump 3 uses too much power in patient circuit (e 21) during a cec.There is no report of any patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER COOLER 3T (230 VOLT)
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80939
GM  80939
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18747434
MDR Text Key336880572
Report Number9611109-2024-00088
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeFR
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Remedial Action Other
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/06/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse