Literature was reviewed regarding the electromagnetic field generated by a left ventricular assist device (vad) on leadless implantable pulse generator (ipg) function.The authors described a patient who was implanted with a leadless ipg that already had a vad implanted from approximately five years prior.Due to the electromagnetic field from the vad, they observed lack of detection, pacing, and undersensing from the leadless ipg.The device was reprogrammed, and it remains in use.No adverse patient effects or additional product performance issues were reported.
|
This information is based entirely on journal literature.Medtronic was made aware of this event through a search of literature publications.This event occurred outside the us.Patient information is limited due to confidentiality concerns.The event only occurred with one patient but specific details on the patient were not provided.Referenced article: impact of the electromagnetic field generated by the left ventricular assist device on leadless pacemaker function.Kardiologia polska.2023.81, 1157¿1159.Doi: 10.33963/v.Kp.97863 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|