Catalog Number 31-301852 |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/31/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).G2: new zealand the device will not be returned for analysis; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch will be submitted.
|
|
Event Description
|
It was reported that during the procedure, the end of the screwdriver sheared off completely.There was no harm or health consequences to the patient.It was reported that no further information is available.
|
|
Event Description
|
No additional event information to report at this time.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Manufacturer Narrative
|
(b)(4).Visual examination of the returned product identified the tip of the device has fractured.There are wear lines on the shaft of the device.The features of the fracture surface visually align with a torsional overload fracture failure mode.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.This complaint was confirmed based on the returned, fracture device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further information is available at the time of this report.
|
|