E2 - initial reporter address 1: (b)(6).Device evaluated by mfr.: the device was returned for analysis.The handshake connection, sheath, coil, burr and annulus were visually examined.Inspection of the device found that the sheath was damaged at the burr housing strain relief.Functional testing was performed with a test advancer.After the returned burr catheter was connected to the test advancer, the advancer was connected to the liquid infusion line.When liquid infusion was turned on, fluid leaked from the damaged portion of the sheath.
|