Catalog Number 382544 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/11/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that bd insyte autog bc has foreign matter on its tip.The following information was provided by the initial reporter: frayed piece of plastic noted on the tip of iv (intra venous) catheter.
|
|
Manufacturer Narrative
|
H.3.A follow up mdr will be submitted if additional information, a device evaluation, or a device history review is completed.E1.Address information was not provided, therefore, xx was used as a place holder.E4.The initial reporter also notified the fda on 11-jan-2024.Medwatch report is mw5150112.
|
|
Manufacturer Narrative
|
Investigation results: no photographic evidence or physical samples were accessible to investigate the reported condition.Our quality engineer team conducted a review of the device history record for the provided material number 382544 and lot number 3248256.This review did not uncover any abnormalities during the production process that could have caused this defect, and all quality tests were found to be within the specified standards.Unfortunately, the exact cause of this incident could not be determined based on the information available.We will continue to track and analyze complaints related to this device and the reported condition in order to identify any emerging trends.
|
|
Event Description
|
No additional information.
|
|