Catalog Number UNKNOWN
Device Problem Detachment of Device or Device Component (2907)
Patient Problems Internal Organ Perforation (1987); Foreign Body In Patient (2687); Device Embedded In Tissue or Plaque (3165)
Event Date 02/01/2024
Event Type  Injury  
Event Description
Description of event according to initial reporter: celect ivc filter found to be multiply fractured with one arm in right pulmonary artery (removed), half of one leg embedded in spine (not removed as was extravascular), and half of one leg retained in filter itself but otherwise loose ie not in wall.Tip embedded.Removed filter and fragment in filter with forceps.Patient outcome: the patient required intervention to prevent permanent impairment/damage.The patient is doing well.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6
bjaeverskov 4632
DA  4632
MDR Report Key18751819
MDR Text Key335933744
Report Number3005580113-2024-00015
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Distributor
Reporter Occupation Physician
Type of Report Initial
Report Date 02/21/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA02/21/2024
Event Location Hospital
Date Report to Manufacturer02/05/2024
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/21/2024
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight76 KG
-
-