Catalog Number 381844 |
Device Problem
Device Contamination with Chemical or Other Material (2944)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/06/2024 |
Event Type
malfunction
|
Event Description
|
It was reported that bd insyte autoguard gn 18ga x 1.16in had foreign matter.The following information was provided by the initial reporter, translated from chinese to english: when the product was opened, the steel needle was found to be black.
|
|
Manufacturer Narrative
|
H.3.A device evaluation and/or device history review is anticipated but is not complete.Upon completion, a supplemental report will be filed.
|
|
Event Description
|
No additional information.
|
|
Manufacturer Narrative
|
Investigation results: the complaint of black material on the needle and catheter was confirmed; however, the source and composition of the material could not be determined from the photograph that was provided for investigation.The photo showed an 18g insyte autoguard iv catheter and needle.A dark colored material was observed on the needle bevel and catheter.The observable features on the submitted photographs did not contain enough information to identify a specific root cause of the reported event.A review of the inspection records and quality/manufacturing controls for the implicated lot indicated no issues with the manufacturing process.Manufacturing controls are in place to mitigate the occurrence of this type of failure.The complaint has been documented and will continue to be monitored as part of ongoing efforts to identify potential manufacturing related issues.
|
|