Model Number V60
Device Problem Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/14/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that while performing preventive maintenance service, the v60 ventilator displayed an alarm for a flow sensor error.There was no patient involvement when the issue occurred.There was no patient or user harm reported.
 
Manufacturer Narrative
A service engineer (se) evaluated the device and reported that the flow and pressure sensors were out of range.The se noted that the gas delivery system (gds) needed to be replaced.The gds was replaced to resolve the reported issue.The flow and pressure sensors were now passing testing per the service manual.The system meets the specification for the performed service and is returned to use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18754097
MDR Text Key335966697
Report Number2518422-2024-09058
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberV60
Device Catalogue Number1053617
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received02/21/2024
Supplement Dates Manufacturer ReceivedNot provided
Supplement Dates FDA Received03/14/2024
Date Device Manufactured01/29/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-