• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: GYRUS ACMI, INC. M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG; HYSTEROSCOPE (AND ACCESSORIES)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

GYRUS ACMI, INC. M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG; HYSTEROSCOPE (AND ACCESSORIES) Back to Search Results
Model Number REXM3-30A
Device Problem Sharp Edges (4013)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was reported, the autoclavable foroblique telescope had burning on the tip, leaving a sharp area.There were no reports of patient harm.
 
Manufacturer Narrative
The device was returned, and the evaluation found the distal tip with a sharp edge.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information a correction to d4.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's investigation and correction to h4.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the root cause of the observed damage on the distal end resulted from improper handling of the device.The device ifu (instructions for use_ 99-1080_bg) states: "study this manual and other labeling thoroughly for safe handling and storage.Misuse of instruments can cause injury to the patient and could have an adverse effect on the procedure being performed.Do not drop instruments or allow them to be struck by other objects;" "keep the distal tip of any electrode, probe, laser fiber, or other ancillary device in the field of view at all times when active." ( page 4) olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M3-GOLD AUTOCLAVABLE FOROBLIQUE TELESCOPE 30 DEG
Type of Device
HYSTEROSCOPE (AND ACCESSORIES)
Manufacturer (Section D)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer (Section G)
GYRUS ACMI, INC.
9600 louisiana avenue north
brooklyn park MN 55445
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18754252
MDR Text Key335966665
Report Number3011050570-2024-00081
Device Sequence Number1
Product Code HIH
UDI-Device Identifier00821925007802
UDI-Public(01)00821925007802(11)220604
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K890328
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberREXM3-30A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/29/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/29/2024
Initial Date FDA Received02/21/2024
Supplement Dates Manufacturer Received03/06/2024
03/25/2024
Supplement Dates FDA Received03/18/2024
03/25/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/04/2022
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-