It was reported that during patient use, the nkv-330 ventilator was alarming circuit disconnect and o2 supply low and went into standby mode.The patient was able to remove the noninvasive facemask and call for assistance.The patient''s oxygen saturation dropped to 83%.The patient recovered quickly and was placed on another ventilator.There was no harm to the patient.A review of the debug logs confirmed that the ventilator had multiple circuit disconnects and o2 supply low alarm activations.When the circuit disconnect alarm is active, the ventilator stops ventilation until it detects a circuit connection or is manually restarted.The o2 supply low alarm occurs when the oxygen supply pressure to the high-pressure oxygen port is less than (b)(4).After this incident was reported to nkom, the hospital checked their oxygen supply lines to patient rooms in this area of the hospital and discovered that the pressure was lower than in other areas.
|