RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 |
Device Problem
Infusion or Flow Problem (2964)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/09/2024 |
Event Type
malfunction
|
Event Description
|
Philips received a complaint from the customer, reporting that the v60 ventilator had the fio2 (fraction of inspired oxygen) setting at 100%, but the spo2 (peripheral capillary oxygen saturation) did not improve.There has been no report of patient or user harm.Further information regarding the reported event has been requested.
|
|
Manufacturer Narrative
|
E1 reporting institution phone number -(b)(6).Reporter phone number - (b)(6).
|
|
Manufacturer Narrative
|
Further details regarding the reported event and the patient could not be obtained despite multiple attempts to retrieve additional information.A service engineer (se) reported that the device has been evaluated.The evaluation did not confirm that the device had a fio2 (fraction of inspired oxygen) issue.There were no error codes observed in the v60 ventilator.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
|
|