Model Number V60
Device Problem Infusion or Flow Problem (2964)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/09/2024
Event Type  malfunction  
Event Description
Philips received a complaint from the customer, reporting that the v60 ventilator had the fio2 (fraction of inspired oxygen) setting at 100%, but the spo2 (peripheral capillary oxygen saturation) did not improve.There has been no report of patient or user harm.Further information regarding the reported event has been requested.
 
Manufacturer Narrative
E1 reporting institution phone number -(b)(6).Reporter phone number - (b)(6).
 
Manufacturer Narrative
Further details regarding the reported event and the patient could not be obtained despite multiple attempts to retrieve additional information.A service engineer (se) reported that the device has been evaluated.The evaluation did not confirm that the device had a fio2 (fraction of inspired oxygen) issue.There were no error codes observed in the v60 ventilator.The complaint will be processed for closure.If additional information becomes available at a later date, the complaint will be reopened, and a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RESPIRONICS
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
1001 murry ridge lane
murrysville, PA 15668
7247330200
MDR Report Key18758125
MDR Text Key336568840
Report Number2518422-2024-09149
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodePO
PMA/PMN Number
K102985
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Device Catalogue Number1053614
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 02/09/2024
Initial Date FDA Received02/22/2024
Supplement Dates Manufacturer Received03/13/2024
Supplement Dates FDA Received04/05/2024
Date Device Manufactured12/04/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-