It was reported this was an off-label mitraclip procedure to treat tricuspid regurgitation (tr) with a grade of 5 and a rotated heart.It was noted imaging was difficult throughout the procedure.An xtw clip was inserted and deployed on the tricuspid valve.However, the clip expulsed and embolized to the hepatic vein.The clip was removed from the anatomy via snare and replaced.The procedure was continued, and two new xtw clips were successfully implanted, reducing tr to a grade of 4.There was no clinically significant delay in the procedure.
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to the reported event.Additionally, a review of the complaint history identified no similar complaints from the lot.All available information was investigated the reported image resolution poor was related to patient and procedural conditions as imaging was reported to be very challenging due to rotated heart.Slda associated with the clip detaching from the septal leaflet appears to be due to the patient conditions (large gaps with tethered leaflets) and therefore, related to off-label use.The reported off-label use of the device was associated with the mitraclip device being used on the tricuspid valve to tr.Unexpected medical intervention was a result of case-specific circumstances.There is no indication of a product issue with respect to manufacture, design or labeling.H6: device code 1494 - indication for use (use in tricuspid valve).
|