• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STAAR SURGICAL COMPANY IMPLANTABLE COLLAMER LENS (ICL); PHAKIC INTRAOCULAR LENS Back to Search Results
Model Number VTICM5_13.2
Device Problems Device Dislodged or Dislocated (2923); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Claim #: (b)(4).
 
Event Description
The reporter indicated that the surgeon implanted an implantable collamer lens into the patient's left eye (os).It was indicated there was rotation without low vaulting.Lens remains implanted.Attempts to obtain additional information have not been successful.If additional information is received a supplemental medwatch report will be submitted.
 
Manufacturer Narrative
Additional data: h6- health effect - impact code (f): "2199" should be removed and "4629" and "4628" should be added.H6- medical device problem code (a): "2993" should be added.H6- type of investigation code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.B5- the reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens of diopter -13.00/3.0/103 (sphere/cylinder/axis) into the patient's left eye (os) (b)(6) 2023.The patient experienced a lens rotation not associated with a low vault.On (b)(6) 2023 the lens was repositioned but that did not resolve the problem.On (b)(6) 2024 the lens was exchanged with a different model, longer length lens and the problem was resolved.The status of the eye was noted to be "all good, patient is happy".The reporter indicated the cause of the event is unknown.Claim# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IMPLANTABLE COLLAMER LENS (ICL)
Type of Device
PHAKIC INTRAOCULAR LENS
Manufacturer (Section D)
STAAR SURGICAL COMPANY
1911 walker avenue
monrovia 91016
Manufacturer (Section G)
STAAR SUGICAL COMPANY
1911 walker avenue
monrovia CA 91016
Manufacturer Contact
joselene muniz
1911 walker avenue
monrovia, CA 91016
MDR Report Key18761119
MDR Text Key336470932
Report Number2023826-2024-00792
Device Sequence Number1
Product Code MTA
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
P030016
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberVTICM5_13.2
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/02/2024
Initial Date FDA Received02/22/2024
Supplement Dates Manufacturer Received02/28/2024
Supplement Dates FDA Received02/28/2024
Date Device Manufactured04/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexMale
-
-