Additional data: h6- health effect - impact code (f): "2199" should be removed and "4629" and "4628" should be added.H6- medical device problem code (a): "2993" should be added.H6- type of investigation code: 4110- lens work order search-no similar complaint event(s) within associated lots were found.B5- the reporter indicated the surgeon implanted a 12.6mm vticm5_12.6 implantable collamer lens of diopter -13.00/3.0/103 (sphere/cylinder/axis) into the patient's left eye (os) (b)(6) 2023.The patient experienced a lens rotation not associated with a low vault.On (b)(6) 2023 the lens was repositioned but that did not resolve the problem.On (b)(6) 2024 the lens was exchanged with a different model, longer length lens and the problem was resolved.The status of the eye was noted to be "all good, patient is happy".The reporter indicated the cause of the event is unknown.Claim# (b)(4).
|