Model Number FV701T
Device Problem Mechanical Problem (1384)
Patient Problem Hydrocephalus (3272)
Event Date 01/26/2024
Event Type  Injury  
Manufacturer Narrative
Investigation: visual inspection: during the visual inspection a deformation of the outer housing of the prosa valve could be determined.The measurement of the plane parallelism could confirm that with a value of - 0.044 mm - outside tolerance (0 ± 0.02 mm).Permeability test: a permeability test has shown that the valve is permeable.Adjustment test: the prosa valve was tested and is not adjustable throughout the normal range.Braking force and brake function test: the brake functionality test has shown that the brake function is operational.However, the braking force cannot be measured due to the non-adjustability of the valve.Internal inspection : after dismantling of the valve, deposits were found in prosa.Results: based on our investigation results, we can determine adjustment difficulties in the valve.The determined deposits can be named as the cause for the adjustment difficulties.Organic material in the cerebrospinal fluid can influence the function temporarily and are known side effects in hydrocephalus therapy.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.
 
Event Description
It was reported that a prosa (#fv701t) was implanted during a procedure.According to the complainant, the valve had adjustment difficulties.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Age: 81 years, height: 180 cm, weight: 80 kg.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROSA VALVE
Type of Device
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
Manufacturer (Section G)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM   14469
Manufacturer Contact
joerg knebel
ulanenweg 2
potsdam, 14469
GM   14469
MDR Report Key18762179
MDR Text Key336052233
Report Number3004721439-2024-00032
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K161853
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2021
Device Model NumberFV701T
Device Catalogue NumberFV701T
Device Lot Number20029092
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/05/2024
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured01/25/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age81 YR
Patient Weight80 KG
-
-