• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALERE SAN DIEGO, INC. FISHER SURE-VUE HCG STAT SERUM/URINE; VISUAL, PREGNANCY HCG, PRESCRIPTION USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALERE SAN DIEGO, INC. FISHER SURE-VUE HCG STAT SERUM/URINE; VISUAL, PREGNANCY HCG, PRESCRIPTION USE Back to Search Results
Model Number FHC-A202
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2017
Event Type  Injury  
Manufacturer Narrative
The date of event was not provided; however, the article indicated the results were obtained between 1sep2017 and 31aug2018.Investigation conclusion: the case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Manufacturing batch record review and retain device testing could not be performed as a lot number was not provided.Review of the risk management report for this product found that the reported issue is within the risk profile for this device; no new hazard has been identified.A root cause could not be determined from the available information as the reported issue was not replicated during testing of retention product.Complaints are tracked and trended on a monthly basis.Per the package insert: very dilute urine specimens, as indicated by a low specific gravity, may not contain representative levels of hcg.If pregnancy is still suspected, a first morning urine specimen should be collected 48 hours later and tested.False negative results may occur when the levels of hcg are below the sensitivity level of the test.When pregnancy is still suspected, a first morning serum or urine specimen should be collected 48 hours later and tested.This test reliably detects intact hcg up to 500,000 miu/ml.It does not reliably detect hcg degradation products, including free-beta hcg and beta core fragments.Quantitative assays used to detect hcg may detect hcg degradation products and therefore may disagree with the results of this rapid test.As with any assay employing mouse antibodies, the possibility exists for interference by human anti-mouse antibodies (hama) in the specimen.Specimens from patients who have received preparations of monoclonal antibodies for diagnosis or therapy may contain hama.Such specimens may cause false positive or false negative results.This test provides a presumptive diagnosis for pregnancy.A confirmed pregnancy diagnosis should only be made by a physician after all clinical and laboratory findings have been evaluated.
 
Event Description
Review of the article "false-negative urine human chorionic gonadotropin testing in the clinical laboratory" noted false negative hcg results for multiple patients while using the sure-vue stat serum/urine hcg test.The eleventh patient mentioned presented with lower abdominal pain.Testing with a urine sample on the sure-vue stat serum/urine hcg test yielded a negative result.The patient had a serum hcg test performed which yielded a positive result of 36 miu/ml on the same day.The patient was believed to have an estimated gestational age of 6 weeks when the negative hcg result was obtained.As a result of the initial negative hcg result, the patient was on medication where the article indicates "data suggests risk during pregnancy" according to the briggs drugs in pregnancy and lactation within the lexicomp online database.No further information was provided.Please see mdr 2027969-2024-00024, 2027969-2024-00025, 2027969-2024-00026, 2027969-2024-00027, 2027969-2024-00028, 2027969-2024-00029, 2027969-2024-00030, 2027969-2024-00031, 2027969-2024-00032, 2027969-2024-00033, and 2027969-2024-00035 for related events.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FISHER SURE-VUE HCG STAT SERUM/URINE
Type of Device
VISUAL, PREGNANCY HCG, PRESCRIPTION USE
Manufacturer (Section D)
ALERE SAN DIEGO, INC.
9942 mesa rim rd
san diego CA 92121
Manufacturer Contact
monika burrell
9942 mesa rim rd
san diego, CA 92121
8588052506
MDR Report Key18763624
MDR Text Key336054681
Report Number2027969-2024-00034
Device Sequence Number1
Product Code JHI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K062361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Literature,Health Professional
Reporter Occupation Physician
Type of Report Initial
Report Date 02/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFHC-A202
Device Catalogue Number23900531
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age47 YR
Patient SexFemale
-
-