Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
Per visual inspection: no physical damage to cartridge holder or inpen front and back shell was noted.Several attempts were made to pair inpen, every time app displayed inpen not found".The inpen does not pair with commercial mobile app.Inpen did not transmit to manufacturing app.Unable to perform baseline wireless/functionality test or displacement doe accuracy test.Inpen passed dialing alignment using dose knob zero alignment gage.Performed battery investigation and measured.0349 v.Unit reached end of life.Unable to confirm dose log inaccuracy due to inpen not communicating to mobile app.In conclusion: inpen failed to communicate to commercial app or manufacturing app.Inpen battery lifetime last 1 year after first paring.It is possible battery life decreased as expected and the low battery icon appears several times near the end of the battery lifetime.Inpen working as designed.Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|