• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA USA, INC. PULSE GEN MODEL 1000; GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA USA, INC. PULSE GEN MODEL 1000; GENERATOR Back to Search Results
Model Number 1000
Device Problems Failure to Deliver Energy (1211); Low impedance (2285); Physical Resistance/Sticking (4012)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Manufacturer Narrative
Livanova usa, inc.Submits this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation, based on information that livanova has obtained, but may not have been able to investigate or verify prior to the date the report was required by the fda.This report does not constitute an admission, or a conclusion by fda or anyone else, that the device, livanova, or livanova's employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects¿ or "malfunctions¿.These words are incorporated into the fda 3500a medwatch form by the fda, and livanova objects to their use.
 
Event Description
It was reported that the patient was seen to have low impedance and was referred for surgery.The patient underwent surgery and only their generator was replaced.The lead impedance was within normal limits after the generator was replaced.Although the impedance resolved with only the generator being changed, there is not enough evidence at this time to rule out a lead issue.No known surgical intervention has occurred to date with the lead.No other relevant information has been received to date.
 
Event Description
Product analysis was completed on the returned generator.An interrogation and system diagnostic tests were performed.The device output signal was monitored for more than 24-hrs and did not perform as expected.A magnet swipe was done to activate the magnet current to capture the duts ramps, so all duts ramps could be captured on the first page.However, the reed switch appears to have stuck closed as the output signal did return after the magnet was applied.Tapped on the dut for the first hour, but the output signal did not return.It appears the reed switch released on its own as the output signal returned approximately 5 hours after the 24-hour monitoring period had begun.A comprehensive automated electrical evaluation (shows an ifi=no condition) was performed.Other than the reed switch stuck closed condition during the output signal monitoring period, the dut performed according to functional specifications.Device history records for the generator was performed.The generator passed final functional and quality specifications prior to release for distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PULSE GEN MODEL 1000
Type of Device
GENERATOR
Manufacturer (Section D)
LIVANOVA USA, INC.
100 cyberonics blvd
houston TX 77058
Manufacturer (Section G)
LIVANOVA USA, INC.
100 cyberonics blvd
suite 600
houston TX 77058
Manufacturer Contact
cindy scott
100 cyberonics blvd
suite 600
houston, TX 77058
2816672681
MDR Report Key18766432
MDR Text Key336081628
Report Number1644487-2024-00182
Device Sequence Number1
Product Code LYJ
UDI-Device Identifier05425025750405
UDI-Public05425025750405
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P970003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model Number1000
Device Lot Number7651
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Event Location Other
Initial Date Manufacturer Received 01/30/2024
Initial Date FDA Received02/23/2024
Supplement Dates Manufacturer Received03/29/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/28/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age15 YR
Patient SexMale
-
-