Model Number BF-1T180 |
Device Problem
Microbial Contamination of Device (2303)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/16/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
|
|
Event Description
|
It was reported that, during reprocessing, the bronchovideoscope's working channel tested positive for coagulase-negative staphylococci and pseudomonas aeruginosa on (b)(6) 2024.The suction channel tested positive for pseudomonas aeruginosa on (b)(6) 2024, and the scope surface tested positive for coagulase-negative staphylococci on (b)(6) 2024.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct d9 a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material could not be identified nor the root cause of it.The events can be detected/prevented in accordance with the following instructions for use: ¿failure to properly clean and high-level disinfect or sterilize endoscopic equipment after each procedure may compromise patient safety.To minimize the risk of transmitting diseases from one patient to another, after each procedure the endoscope and its ancillary equipment must undergo thorough manual cleaning followed by high-level disinfection or sterilization as described in chapter 3, ¿cleaning, disinfection, and sterilization procedures¿.Reprocess not only the external surface of the endoscope but also all channels.¿ olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|