• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AIZU OLYMPUS CO., LTD. EVIS EXERA II BRONCHOVIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AIZU OLYMPUS CO., LTD. EVIS EXERA II BRONCHOVIDEOSCOPE Back to Search Results
Model Number BF-1T180
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/16/2024
Event Type  malfunction  
Manufacturer Narrative
The suspect device has not yet been returned to olympus for evaluation and the investigation is ongoing.The physical device evaluation has not been completed.Prior to the device evaluation, the device was sent out for additional microbiological testing.The microbiological analysis results are pending.Once the investigation has been completed, a supplemental report will be submitted with device evaluation results.
 
Event Description
It was reported that, during reprocessing, the bronchovideoscope's working channel tested positive for coagulase-negative staphylococci and pseudomonas aeruginosa on (b)(6) 2024.The suction channel tested positive for pseudomonas aeruginosa on (b)(6) 2024, and the scope surface tested positive for coagulase-negative staphylococci on (b)(6) 2024.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and to correct d9 a review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the foreign material could not be identified nor the root cause of it.The events can be detected/prevented in accordance with the following instructions for use: ¿failure to properly clean and high-level disinfect or sterilize endoscopic equipment after each procedure may compromise patient safety.To minimize the risk of transmitting diseases from one patient to another, after each procedure the endoscope and its ancillary equipment must undergo thorough manual cleaning followed by high-level disinfection or sterilization as described in chapter 3, ¿cleaning, disinfection, and sterilization procedures¿.Reprocess not only the external surface of the endoscope but also all channels.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II BRONCHOVIDEOSCOPE
Type of Device
BRONCHOVIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18767095
MDR Text Key337078546
Report Number9610595-2024-03828
Device Sequence Number1
Product Code EOQ
Combination Product (y/n)N
Reporter Country CodeHU
PMA/PMN Number
K061313
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 04/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBF-1T180
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received02/23/2024
Supplement Dates Manufacturer Received04/23/2024
Supplement Dates FDA Received04/23/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/26/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-