Edwards received notification of a pascal in mitral procedure where the guide sheath was aspirating air bubbles after insertion of the implant past seals and loader removed.It was verified that the device was off anatomy and even made small movements to the guide adding flex and retracting as well as advancing.The guide sheath was removed, and new guide sheath prepped and used with the same issue- there was air while aspirating.Same mitigation techniques were used but another sheath was unavailable, so the procedure was continued as normal.
|
Additional information received from the clinical specialist stated that there were no ekg changes and there were no issues during preparation.There were no device issues during or post implantation as well.Per case submission la pressure started at 22 and dropped to 10 post procedure.The patient was under general anesthesia, but no air was ever introduced into the patient.Also, the patient is doing fine and there were no consequences.
|
The following sections were updated/corrected/added: b4, b5, g3, g6, h2 and h10.
|
The following sections were updated/corrected/added: b4, d4, g3, g6, h2, h3, h4, h6 and h11 the complaint for inadequate aspiration, air remaining in device during insertion was confirmed with objective evidence.The complaint was able to be recreated but after drying the unit, a burst of air bubbles during aspiration was observed.A minor leak was found emerging from the guide sheath pull wire, however it was not believed to be a significant contributing factor to the event as the leak was only seen under high pressure.Visual evidence of spine to shaft bond profile change before and after drying the unit points to delamination as a probable root cause.Based on the investigation, delamination of the spine to shaft bond resulted in the leak path allowing air to continuously enter the system during aspiration as detailed in the event description.This delamination can occur due to a combination of factors, however, a definite root cause is unable to be determined.Furthermore, the device history record review was completed, and this device passed all manufacturing and sterilization inspections.No nonconformances related to the complaint event were identified.
|