During follow-up, loss of capture and r wave amplitude variation was observed on the right ventricular (rv) lead.During revision, the helix of the rv lead did not retract or extend.The rv lead was explanted and replaced to resolve the event.The patient was stable and there were no adverse consequences.
|
The reported events were failure to capture, r-wave amplitude variation and helix mechanism issue.As received, a complete lead was returned in one piece.The reported event of helix mechanism issue was confirmed.Visual examination of the lead found the helix was extended and clogged with blood and tissue.X-ray examination of the lead found the inner coil in the connector region was over torqued due to procedural damage.After cutting the lead, cleaning the distal portion of the lead, and applying torque directly to the inner coil, the helix was able to extend and retract.The measured full helix extension length was within specification.The reported event of helix mechanism issue was isolated to the over-torqued inner coil in the connector region and the helix clogged with blood and tissue.The reported events of failure to capture and r-wave amplitude variation were not confirmed.Visual and x-ray examination of the lead did not find any anomalies with the exception of procedural damage.Electrical testing of the lead did not find any indication of conductor fractures or internal shorts.
|