• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT JOT DX¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT JOT DX¿ ICM; RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION) Back to Search Results
Model Number DM4500
Device Problem Failure to Interrogate (1332)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/11/2024
Event Type  malfunction  
Event Description
During an implant procedure, the device was unable to be interrogated.The device was explanted and replaced to resolve the event.The patient was stable and will continue to be monitored.
 
Manufacturer Narrative
The customer complaint of interrogation problem was not confirmed.The device was received in normal working conditions with the battery voltage in normal range of operation.Analysis was performed, including telemetry and electrical testing which indicated normal device functionality.A longevity assessment was performed and the device was in the normal range of operation with appropriate remaining longevity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JOT DX¿ ICM
Type of Device
RECORDER, EVENT, IMPLANTABLE CARDIAC, (WITHOUT ARRHYTHMIA DETECTION)
Manufacturer (Section D)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ABBOTT
15900 valley view court
sylmar CA 91342
Manufacturer Contact
richard williamson
15900 valley view court
sylmar, CA 91342
MDR Report Key18768692
MDR Text Key336122856
Report Number2017865-2024-33342
Device Sequence Number1
Product Code MXC
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K163407
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberDM4500
Device Lot NumberP000183481
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/08/2024
Initial Date FDA Received02/23/2024
Supplement Dates Manufacturer Received03/29/2024
Supplement Dates FDA Received04/10/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/14/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-