Product complaint # (b)(4).E3 initial reporter occupation: lawyer.H6 component code: appropriate term/code not available (g07002) used to capture no findings available.Investigation summary : no device associated with this report was received for examination.The product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A records evaluation (mre) was not perform.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
Update ad 15 feb 2024 medical records received.Patient underwent revision due to elevated cobalt levels- removal of cobalt chrome hip ball and cobalt chrome acetabular liner.Slightly cloudy yellow synovial fluid in moderate quantity, associated with reactive, tan-colored synovitis consistent with metallosis, trunnion corrosion identified.This included small thin pieces of dark stained material adherent to the base of the trunnion.Periprosthetic osteolysis and wear of articular bearing surface.For placement with highly cross-linked polyethylene liner and revision of ceramic hip ball.Doi: (b)(6) 2009, dor: (b)(6) 2021.Affected side : left hip.This is a bilateral hip arthroplasty.A link of (b)(4).
|