Catalog Number 212760 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 02/01/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
As part of the normal complaint follow-up, an evaluation of the event has been initiated by mako surgical.A supplemental report will be submitted if additional information becomes available.
|
|
Event Description
|
Screw fell out of offset reamer in spd.Found screw and is with reamer.Screw is not in patient.Case type / application: tha 4.1.
|
|
Manufacturer Narrative
|
An event regarding disassociation involving a mako reamer handle was reported.The event was confirmed method & results: product evaluation and results: functional inspection confirms the offset reamer screw is loose.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been other complaints with similar event(s) for the lot referenced.Conclusions: the alleged failure mode was confirmed.No additional investigation or specific actions are required.If additional information is received then the complaint will be reopened.
|
|
Event Description
|
Screw fell out of offset reamer in spd.Found screw and is with reamer.Screw is not in patient.Case type / application: tha 4.1.
|
|