MAUDE Adverse Event Report: MPRI ATTAIN STARFIX LEAD; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ATTAIN STARFIX LEAD; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P Back to Search Results
Model Number 419588
Device Problems High impedance (1291); High Capture Threshold (3266)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/25/2024
Event Type  malfunction  
Event Description
It was reported that one day post implant the patient was feeling pain on their side and heard an audible noise from their cardiac resynchronization therapy defibrillator (crt-d).Upon review it was noted that there were several right ventricular (rv) lead impedance alerts including for the pacing, svc defibrillation and rv defibrillation coils.It was also noted that there were high left ventricular (lv) lead thresholds and lv impedance was high.The patient was brought back to the lab and it was determined that the rv lead was not properly seated in the crt-d header.The pocket was reopened and the lead was appropriately connected to the device. no further patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Continuation of d10:  5867-3m adaptor implanted: (b)(6) 2024 693565 lead implanted: (b)(6) 2024 etlw1620c156e stent graft implanted (b)(6) 2015 esbf2814c103e stent graft implanted (b)(6) 2015 etlw1616c156e stent graft implanted (b)(6) 2015 5076-52 lead implanted: (b)(6) 2010 medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN STARFIX LEAD
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) P
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key18771357
MDR Text Key336333524
Report Number2649622-2024-05182
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P060039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/23/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date10/17/2010
Device Model Number419588
Device Catalogue Number419588
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 01/25/2024
Initial Date FDA Received02/23/2024
Date Device Manufactured09/24/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
6719 ADAPTOR, DTPB2D1 CRT-D, 5867AS ADAPTOR
Patient Age76 YR
Patient SexMale
Patient Weight98 KG
-
-