• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SHIRAKAWA OLYMPUS CO., LTD. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Device Problems Thermal Decomposition of Device (1071); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/03/2024
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and the evaluation found the lamp wire was worn and there was a power unit failure.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported, the light source lamp wire was worn and the power supply voltage was unstable.The issue occurred during preparation for use on a bronchoscope diagnostic procedure.The facility completed this procedure using a similar device.There were no reports of delay or patient harm associated with the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HALOGEN LIGHT SOURCE
Type of Device
HALOGEN LIGHT SOURCE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA   961-8061
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18778013
MDR Text Key336215588
Report Number3002808148-2024-01755
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/04/2024
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/19/2024
Initial Date FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CHF-P60
-
-