COLOPLAST A/S ALTIS SINGLE INCISION SLING SYSTEM; FEMALE STRESS URINARY INCONTINENCE SURGICAL MESH-SLING, SYNTHETIC POLYMER
|
Back to Search Results |
|
Catalog Number 519650 |
Device Problem
Difficult or Delayed Separation (4044)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/29/2023 |
Event Type
malfunction
|
Event Description
|
According to the available information altis sling was opened in the operation room.With the two helical trocars, doctor successfully implanted both anchors of the first sling.However, when he went to tension the sling with the tensioning suture, the dynamic anchor separated from the suture.Consequently, he cut the other side of the sling and removed the product.Another altis sling was opened and implanted successfully.There were no issues reported with the patient.
|
|
Manufacturer Narrative
|
The lot number was reviewed for complaint trend, nonconforming report, and capa.No ncs nor capas were identified in veeva.There were several complaints identified against this lot; however, with various failure modes.Therefore, this is not considered a trend.Devices met specification prior to release.
|
|