MAUDE Adverse Event Report: AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AOMORI OLYMPUS CO., LTD. SINGLE USE ELECTROSURGICAL KNIFE Back to Search Results
Model Number KD-611L
Device Problems Break (1069); Crack (1135); Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/27/2022
Event Type  Injury  
Manufacturer Narrative
During device evaluation at olympus, it was found the tip was cracked and part of the tip had detached from the electrode.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over one year since the subject device was manufactured.Although a definitive root cause of the tip detachment could not be determined, likely factors causing the defect could have been the following: 1.Due to the factor described below, spark discharge between the tissue and the electrode occurred during output activation.The high-frequency output was set too high.The output setting for activation was too high.The electrosurgical unit was used in the coagulation mode.The output activation time was too long.2.High heat might have been locally applied to the electrodes and tip.The tip melted and cracks developed.The fixing strength of the adhesive securing the chip decreased.3.A force to perform tissue incision was applied to the cracked area, causing the tip to crack and detach.The customer may be able to reduce and prevent occurrence of the event by handling device in accordance with the following ifu : "- when the electrosurgical unit is used in the coagulation mode, crack/detachment of the tip and the electrode or deformation/break of the cutting knife could occur, for example when the high-frequency output is set too high or the length of the contact between the cutting knife and tissue is too short.During treatment, always ensure that the slider slides on the handle smoothly and that the electrosurgical knife observed in the endoscopic image is normal.Should cracks or detachment of the tip or deformation/break of the cutting knife be detected during use, immediately shut off the power supply, discontinue the procedure, pull the slider and withdraw the endoscope from the patient with the cutting knife retreated in the insertion portion of this instrument.Do not continue using an abnormal electrosurgical knife to prevent perforation or hemorrhages.If the tip or cutting knife is detached be sure to collect it using grasping forceps.Aspirate fluids such as mucus that adhere to the electrode and/or the cutting knife, outer sheath and body cavity tissues.Patient injury such as punctures, hemorrhages, mucous membrane damage and thermal injury of tissue could result if output is activated when in contact with these adhering fluids.When current is discharged while the cutting knife is being separated from the mucosa under wet situation, it may break the cutting knife or crack the tip.Always operate the electrosurgical unit at the minimum output level and for the minimum time necessary to successfully complete the procedures.Excessive output level and time may result in patient injury, such as punctures, hemorrhages or mucous membrane damage.Application of high-voltage waveforms for extended periods increase the likelihood that it may break the cutting knife or crack the tip.When a high-voltage waveform has to be used, minimize the duration of current application." if additional information becomes available at a later date, this report will be supplemented.Olympus will continue to monitor the field performance of this device.
 
Event Description
It was reported during a therapeutic endoscopic submucosal dissection (esd), when the doctor was dissecting the tissue, the tip of the single use electrosurgical knife was broken (cracked) and dropped into the patient's body.The debris was collected.The doctor replaced the device and completed the procedure.There was no reported patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE ELECTROSURGICAL KNIFE
Type of Device
SINGLE USE ELECTROSURGICAL KNIFE
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA   036-0357
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18784339
MDR Text Key336294825
Report Number9614641-2024-00479
Device Sequence Number1
Product Code KNS
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K092309
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberKD-611L
Device Lot Number25K
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/08/2022
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/23/2022
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-