• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CORDIS CORPORATION SUPER TORQUE MB; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CORDIS CORPORATION SUPER TORQUE MB; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Catalog Number 532598C
Device Problem Structural Problem (2506)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/30/2024
Event Type  malfunction  
Event Description
As reported, the calibrated markers (1 cm intervals) on a 5f 65cm 8 side-hole (sh) pigtail super torque angiographic catheter had slipped and moved to the base of the catheter.There were no reports of patient injury.This issue was discovered on screen while performing an iliac stenting during an endovascular aneurysm repair (evar).The device will not be returned for evaluation.
 
Manufacturer Narrative
Additional information is pending and will be submitted within 30 days upon receipt.
 
Manufacturer Narrative
As reported, the calibrated markers (1 cm intervals) on a 5f 65cm 8 side-hole (sh) pigtail super torque angiographic catheter had slipped and moved to the base of the catheter.There were no reports of patient injury.This issue was discovered on screen while performing an iliac stenting during an endovascular aneurysm repair (evar).The device was not returned for evaluation.Without the return of the device or images for analysis, the reported customer event ¿marker band (super torque mb_-offset/out of position-prior to/during measurement¿ could not be confirmed.Failure to follow instructions may have contributed to the reported event as this device is contraindicated for use in evar procedures.Users are trained to inspect for signs of damage prior to and during use.Any product with damage is not to be used.Information for safety is provided in the products labeling with the intent to make the user aware of the risks.According to the instructions for use (ifu), although not intended as a mitigation of risk, ¿contraindications: do not use the super torque mb catheter in procedures where entrapment of the catheter between endovascular devices and the vessel wall may occur, for example endovascular aortic repair (evar) procedures.Manipulation of the catheter under excessive friction due to interaction with other devices or while trapped in the vasculature, can lead to stretching or elongation of the catheter.Stretching or elongation of the catheter during endovascular procedures could result in the marker bands moving along the catheter.In extreme cases, marker bands may come off the catheter and dislodge into the vascular system.¿ based on the available information, there is no indication that the event is related to the device design or manufacturing process.Therefore, no preventative or corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPER TORQUE MB
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
CORDIS CORPORATION
14021 nw 60 avenue
miami lakes, florida 33014
Manufacturer (Section G)
CORDIS US CORP.
14201 nw 60 avenue
miami lakes FL 33014
Manufacturer Contact
karla castro
santiago troncoso 808
juarez, chihuahua 
7863138372
MDR Report Key18785527
MDR Text Key336453132
Report Number9616099-2024-00064
Device Sequence Number1
Product Code DQO
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K915836
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 03/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number532598C
Device Lot Number18238085
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 01/31/2024
Initial Date FDA Received02/26/2024
Supplement Dates Manufacturer Received02/29/2024
Supplement Dates FDA Received03/06/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-