• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. M2A-38 CUP NON FLARED SZ 50MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. M2A-38 CUP NON FLARED SZ 50MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Material Erosion (1214)
Patient Problem Metal Related Pathology (4530)
Event Date 09/19/2022
Event Type  Injury  
Event Description
It was reported that approximately thirteen years post-implantation, the patient had a right hip revision due to elevated metal ion levels and concern for metal on metal wear.During the procedure, the patient had a good deal of serosanguineous fluid in the joint and metal pathology was noted.A new liner and head were placed.The stem and cup remained implanted.Attempts have been made and no further information has been provided.
 
Manufacturer Narrative
(b)(4).D10: cat# 11-103203 lot# 553970 taperloc por lat fmrl 9x137.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2024 - 00530.The customer has indicated that the product will not be returned to zimmer biomet for investigation as it remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Review of the device history records identified no deviations or anomalies during manufacturing.Medical records were provided; review of the available records identified the following: an initial right tha was performed.The patient developed elevated metal ions, and a revision was performed.Staining was found on the tissues and fluid found around the joint from a metal on metal reaction.Wear was noted.The head was explanted and replaced.A definitive root cause cannot be determined.This complaint was confirmed based on the provided medical records.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further information is available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A-38 CUP NON FLARED SZ 50MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18787772
MDR Text Key336320596
Report Number0001825034-2024-00531
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011110
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 02/27/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date07/31/2019
Device Model NumberN/A
Device Catalogue Number15-106050
Device Lot Number922530
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/05/2024
Initial Date FDA Received02/27/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/01/2009
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention; Hospitalization;
Patient SexFemale
Patient Weight98 KG
-
-