MEDOS INTERNATIONAL SARL TRUESPAN 24 DEGREE PLGA; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 228162
Device Problem Device-Device Incompatibility (2919)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/13/2024
Event Type  malfunction  
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Device was used for treatment, not diagnosis.H10 additional narrative: e3: reporter is a j&j sales representative.D4: the expiration date is currently unavailable.H4: the device manufacture date is unknown.(b)(4) as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.
 
Event Description
It was reported by the health authority from switzerland that during a knee arthroscopy surgical procedure with knee repair performed on (b)(6) 2024 it was observed that the truespan 24 degree plga device would not fire the back stops.It was reported that the instrument was blocked.It was not reported if a spare device was available for use.It is unknown how the procedure was completed.There was no surgical delay or patient consequences.No additional information was provided.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.D9, h3, h6: the actual device has been returned and is currently pending evaluation.Once reliability engineering evaluates the device, a supplemental medwatch report will be sent accordingly.
 
Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: d4, h4: the expiration date and device manufacture date were reported as unknown on the initial report; and have been updated accordingly.Therefore, udi: (b)(4).Investigation summary the product was returned to depuy synthes mitek for evaluation.The depuy synthes mitek team conducted a visual inspection of the returned device.Visual inspection found that truespan 24 degree plga had the first plate deployed, the plates and the suture were still on the shaft, also the red trigger was broken and completely loose.No other anomalies could be observed.A functional test was unable to be performed due to the condition of the device.The overall complaint was confirmed as the observed condition of the truespan 24 degree plgawould contribute to the complained device issue.Based on the investigation findings, the potential cause could be traced to the procedural variables, such handling of the device or product interaction during procedure, when not inserting the needle to the proper depth for deployment which could have caused blocking insertion, when this occurred resistance may have been felt and the trigger did not pull all the way.As per instructions for use (ifu), it is necessary to use a calibrated probe, measure the width of the meniscal tissue to be repaired and set the adjustable depth.Also, it is important to fully squeeze the red deployment trigger while maintaining depth positioning to deliver the first implant.The implant is fully deployed when you hear an audible ¿click¿.Fully release the trigger after deployment, and it has been determined that no corrective and/or preventative action is proposed.There is no indication that a design or manufacturing issue has caused the reported complaint condition.As part of depuy synthes mitek quality process, all devices are manufactured, inspected, and released to approved specifications.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRUESPAN 24 DEGREE PLGA
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INTERNATIONAL SARL
chemin blanc 38
le locle CH-24 00
SZ   CH-2400
Manufacturer Contact
kate karberg
325 paramount drive
raynham, MA 02767
3035526892
MDR Report Key18788782
MDR Text Key336400652
Report Number1221934-2024-00576
Device Sequence Number1
Product Code MBI
UDI-Device Identifier10886705026050
UDI-Public10886705026050
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K153667
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number228162
Device Lot Number203L317
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/14/2024
Initial Date FDA Received02/27/2024
Supplement Dates Manufacturer Received03/25/2024
04/17/2024
04/17/2024
Supplement Dates FDA Received03/26/2024
04/17/2024
04/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/31/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-