It was reported that during a knee surgery, when the surgeon was securing the meniscus in the jaws of the firstpass mini and twisted it slightly, the device lost its catching blade, one side of the section that catches the mini-tape as it passes through.The metal part was retrieved from the patient.The procedure was completed using a back-up device.There was a delay less than 30 minutes and no further complications were reported.
|
H10: h3, h6: the reported device was received for evaluation.A visual inspection revealed that the device was not returned with any original packaging.One side of the suture capture has been disconnected from the upper jaw.Bio debris is present.No other visual deficiencies were found.A functional evaluation was performed on the returned device and found that pulling the lever will close the jaw and deploy the suture passer needle.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause has been associated with unintended use of the device.Factors that could have contributed to the failure include: (1) excessive force (2) tissue thickness (3) damage or debris on the device tip between passes.No containment or corrective actions are recommended at this time.
|