A user facility returned the olympus loaner, bronchofibervideoscope, to the olympus service center.Before returning the scope, it was processed and microbial test, the results was positive, and the device was isolated.It was reprocessed and retested and the result came back negative.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation and device evaluation.A review of device history record and historical complaints analysis was conducted and no deviations that could have caused or contributed to the reported issue were identified.Based on the results of the investigation, a definitive root cause the observed b30 error/no image issue could not be determined, however, due to an observed pinhole leak in the bending section rubber, it is likely that the issue was due to an ingress of water into the scope connector during user handling, resulting in an electronic component malfunction.The event can be detected/prevented by following the instructions for use (ifu) which states: important information ¿ please read before use precautions: caution turn the video system center on only when the endoscope connector is connected to the light source.In particular, confirm that the video system center is off before connecting or disconnecting the endoscope connector.Failure to do so can result in equipment damage, including destruction of the image sensor.Precautions for disappeared or frozen endoscopic image: warning follow the precautions given below.Otherwise, the endoscopic image may disappear unexpectedly or the frozen image may not be restored during the examination.3.8 inspection of the endoscopic system inspection of the endoscopic image confirm that the wli endoscopic image is normal.5.1 troubleshooting if any irregularity is observed during the inspection described in chapter 3, ¿preparation and inspection¿, do not use the endoscope and solve the problem as described in section 5.2,¿troubleshooting guide¿.If the problem still cannot be resolved, send the endoscope to olympus for repair as described in section 5.4, ¿returning the endoscope for repair¿.Also, should any irregularity be observed while using the endoscope, stop using it immediately and withdraw the endoscope from the patient as described section 5.3, ¿withdrawal of the endoscope with an irregularity¿.Olympus will continue to monitor field performance for this device.
|