Model Number GF-UCT180
Device Problems Material Separation (1562); Failure to Clean Adequately (4048)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Event Description
It was observed that during the device evaluation, the gastrovideoscope exhibited a stain that entered into the lens from gap of adhesive on the distal end.The procedure is unknown.There were no reports of patient involvement.
 
Manufacturer Narrative
The device was returned and the evaluation found the following: scope body has a crack; there is a gap between acoustic lens and ultrasound probe; adhesive on bending section cover or distal sheath rubber has wear; connecting tube has a scratch; and due to wear of angle wire, bending angle in down direction does not meet the standard value.Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation the gap between acoustic lens and ultrasound probe was confirmed.The cause of the gap between acoustic lens and ultrasound probe was not established.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: "do not touch the electrical contacts inside the videoscope cable connector.Ccd damage may result.Do not apply shock to the distal end of the insertion section, particularly the ultrasound transducer and the objective lens surface at the distal end.Visual abnormalities may result.Do not hold the ultrasonic transducer when holding the insertion tube.The ultrasonic transducer damage can result and/or the ultrasonic image will be abnormal.Do not twist or bend the bending section with your hands.Equipment damage may result.Do not squeeze the bending section forcefully.The covering of the bending section may stretch or break and cause water leaks.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS EXERA II ULTRASOUND GASTROVIDEOSCOPE
Type of Device
ULTRASOUND GASTROVIDEOSCOPE
Manufacturer (Section D)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima 961-8 061
JA  961-8061
Manufacturer (Section G)
SHIRAKAWA OLYMPUS CO., LTD.
3-1 okamiyama
odakura, nishigo-mura,
nishishirakawa-gun, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18790045
MDR Text Key336351559
Report Number3002808148-2024-01824
Device Sequence Number1
Product Code ODG
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K093395
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 03/29/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberGF-UCT180
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/11/2024
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 02/13/2024
Initial Date FDA Received02/27/2024
Supplement Dates Manufacturer Received03/21/2024
Supplement Dates FDA Received03/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/26/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-