Model Number 419688 |
Device Problems
Failure to Capture (1081); High impedance (1291); Adverse Event Without Identified Device or Use Problem (2993); High Capture Threshold (3266)
|
Patient Problem
Undesired Nerve Stimulation (1980)
|
Event Date 02/04/2024 |
Event Type
malfunction
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during the implant procedure, the physician tried to place the left ventricular (lv) lead into the target vein but the test parameters showed high thresholds, no pacing, and high impedance.After pulling back a few markers and testing again, the thresholds and impedance remained high and still no pacing.The physician withdrew from the first target vessel, and after much effort, the lead entered the second target vessel.It was noted the threshold of the second blood vessel was also unsatisfactory and there was diaphragmatic bear.The physician returned to the first target vessel and several tests still showed high threshold and slightly high impedance.Good parameters were unable to be obtained so left ventricular pacing was abandoned and switched to left bundle branch pacing.The lead was discarded and a new lead was used.No further patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Medtronic submits this report to comply with fda regulations 21 cfr parts 4 and 803.Medtronic has made reasonable efforts to provide as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.Medtronic will submit a supplemental report if additional relevant information becomes known.
|
|
Search Alerts/Recalls
|
|